Double Beam Spectrophotometer

The DIDAC introduces the first Double Beam Spectrophotometer with a variable spectral bandwidth into the range. The highly stable optics and two detectors measure the sample and reference simultaneously optimising measurement accuracy. The Double Beam Spectrophotometer has measurement modes for photometrics, concentration, multi-wavelength, spectrum scanning, kinetics, quantitation, DNA/RNA and protein analysis.

Specification of Double Beam Spectrophotometer

Wavelength range

190 to 1100nm

Wavelength resolution

0.1nm

Wavelength accuracy

± 0.3nm (at 0.5 and 1nm bandwidth) ± 0.5nm (at 2, 4 and 5nm bandwidth)

Wavelength reproducibility

± 0.2nm

Spectral bandwidth

Variable 0.5, 1, 2, 4, 5nm

Photometric range

- 0.3 to 3.0A 0 to 200%T

Photometric accuracy

± 0.002A (0- 0.5A) ± 0.3% T (0- 100%T)

Photometric reproducibility

± 0.001 Abs (0 to 0.5 Abs) ± 0.002 Abs (0.5 to 1.0 Abs) 0.15%T (0- 100%T)

Resolution

0.1%T, 0.001A

Stray light

< 0.05%T at 220 and 360nm

Noise

0.0005A

Stability

± 0.001A/h at 500nm after 15 min warm up

 


DIDAC INTERNATIONAL

Contact Us

( *represents compulsory fields )

Tips on getting accurate quotes. Please include product name, order quantity, usage, special requests if any in your inquiry.



Contact Details
Regd. Office Address : 507, Shahpuri Tower, C- 58, Janakpuri, New Delhi - 110058, India
Mr. A. Kumar (MD)
Factory: Didac Mension, Phase 1, Mangolpuri Industrial Area, New Delhi-83 INDIA 0091 78381 16798
Trust Stamp
Product Showcase
Quick Enquiry Form
Send Inquiry Send SMS    Call Me Free