Home | » Products | » Lifescience Microscopes | » Confocal Laser Scanning Microscope |
Description of Confocal Laser Scanning Microscope
Confocal Laser Scanning Microscope is designed to meet some of the most difficult challenges in modern science. With the high sensitivity and speed required for live cell and tissue imaging, the also provides an intuitive and adaptable interface, and is capable of microplate imaging and complex screening protocols. The series supports complete workflows from live cell imaging through image processing, like deconvolution, and analysis.
Specification of Confocal Laser Scanning Microscope
Main Laser Combiner |
Violet/Visible Light Laser |
405 nm: 50 mW, 488 nm: 20 mW, 561 nm: 20 mW, 640 nm: 40 mW One optional laser port for sub laser combiner or optional laser unit |
|
Optional Laser |
Sub Laser Combiner |
Laser as follows (max. 3 laser units) 445 nm: 75 mW, 514 nm: 40 mW, 594 nm: 20 mW, fiber connected to main laser combiner |
|
Single Laser Unit |
445 nm: 75 mW, 514 nm: 40 mW, or 594 nm: 20 mW, directly connected to main laser combiner |
||
Laser Light Control |
Main laser combiner with implemented AOTF system, ultra-fast intensity modulation with individual laser lines, additional shutter control continuously variable (0.1 % - 100 %, 0.1 % increments) |
||
Scanner |
Scanning Method |
2 silver-coated galvanometer scanning mirrors |
2 silver-coated galvanometer scanning mirrors |
Pinhole |
Single motorized pinhole, pinhole diameter ÃÆÃ¸50 - 800 Ãâõm (1 Ãâõm Steps) |
||
Field Number (FN) |
18 |
||
Dichromatic Mirror Turret |
8 positions (high performance DMs and 10/90 mirror) |
||
Optional Unit for Scanner |
Laser Power monitor, optional laser port |
||
High Sensitivity-Spectral Detector |
Detector Module |
Cooled GaAsP photomultiplier, 2 channels |
|
Spectral Method |
Motorized Volume Phase Holographic transmission diffraction grating, motorized adjustable slit, |
||
Dichromatic Mirror Turret |
8 positions (high performance DMs and mirror) |
||
Microscope |
Motorized Microscope |
Inverted IX83 (IX83P2ZF) Integrated motorized focus module, minimum increment 0.01 ̵̉m |
|
Fluorescence Illumination Unit |
External fluorescence light source, fiber adapter to optical port of scan unit, motorized switching between LSM light path and fluorescence illumination |
||
Transmitted Light Detector Unit |
Module with integrated external transmitted light photomultiplier detector and LED lamp, motorized switching |