Back to top
Home » Products » Laboratory System And Analyser » Microwave Digestion System

Microwave Digestion System

Microwave Digestion System
Microwave Digestion System
Send Inquiry
Product Code : DILSA070
Brand Name : DIDAC
Product Description

Microwave Digestion System

Microwave Digestion System is a new developed microwave digestion system with all vessels temperature and pressure control. With IR temperature measurement method, our digestion system can realize real-time temperature monitoring of each individual sample. With special mechanic design, the consumable free microwave digestion system is of safety and friendly operation and durability.  

Features:

  • 60L industrial grade chamber is made of 316 stainless steel with PFA coating to prevent emissions and corrosions.
  • Intelligent protection system will alarm and self protect when temperature and pressure turn to abnormal conditions which ensures operation safety.
  • Contact free temperature and pressure measurement which avoids cross-contamination.
  • Outer vessel made of PEEK glass which prevents corrosion and high temperature is high quality material which assures operation safety
  • The system is design to avoid of leakage after over-pressure release which saves batch samples and solutions so as to reduce cost.
  • The system is design to avoid of leakage after over-pressure release which saves batch samples and solutions so as to reduce cost.
  • The system releases pressure vertically to ensure safety during operation.
  • The continuous 360°rotation design assures even digestion and makes system shelf life longer.
  • lid locking design with multiple protection greatly improves uses ease of handing.
  • Lid locking design with multiple protection greatly improves users ease of handling.
  • 12 vessels/batch digestion capacity improves sample preparation efficiency. Individual frame for each vessel makes operation convenient and meets different application requirements.
  • Multi safety protection functions double door prevention, temperature monitoring, high temperature prevention, pressure monitoring, high pressure prevention, instruments abnormal sound monitoring, and lid locking prevention.
  • No consumables are necessary for the microwave digestion system for use which makes uses more convenient..

Specification:

Chamber

Round  Oven Chamber Design

Sample

100ml

Pressure Control

Full Scanning

Pressure Measurement

Touch-free

Temperature Control

Full Scanning

Temperature Measurement

IR

Vessel Material

Inner Vessel Imported TFM Outer Vessel Imported PEEK+ glass fiber

Material of Explosion-Proof Frame

Polypropylene Glass Fiber

Maximum Working Pressure

10 MPa

Maximum Working Temperature

250

Temperature Limit

300

Chamber Volume

60ml

Display

8 Inch Touch Screen

Pressure Control Range

0-6 MPa (0-60kg/cm2)

Pressure Control Accuracy

±0.01MPa(0.1kg/cm2)

Microwave Power

2000W †0~2000 adjustable °

Microwave Frequency

2450MHz

Rotation Mode

360 ° continuous rotation

Microwave Leakage

<5mW/ cm2

Power

AC 220V 10 50/60Hz

Size

640mm x 630mm x 590mm

Weight

75kg



DIDAC INTERNATIONAL
X


Contact Details
Regd. Office Address : 507, Shahpuri Tower, C- 58, Janakpuri, New Delhi - 110058, India
Phone :+918068096079
Mr. A. Kumar (MD)
Mobile :+918068096079
Factory: Didac Mension, Phase 1, Mangolpuri Industrial Area, New Delhi-83 INDIA 0091 78381 16798
Trust Stamp
Product Showcase
Quick Enquiry Form
Send Inquiry Send SMS    Call Me Free